CO2 GASCROMATOGRAPHIC ANALYSIS SYSTEM
General Description
The Carbon Dioxide purity for beverage and food uses, must be compliant to FOOD and ISBT specifications.
Carbon dioxide production need to be continously checked for the determination of total purity and for a lot of single compounds, like sulphur compounds, hydrocarbons, aromatic hydrocarbons, alogenated hydrocarbons, carbon monoxide and permanent gases.
The TPI-Pentatech analyser estabilishing the best design with minimum cost for comply the CO2 specifications.
The TPI-Pentatech system could be combined with multiple sampling system.
Technical Features
Description:
GC based analytical system for pure Carbon Dioxide.
Detectors:
Flame Photometric Detector
Thermal Conductivity Detector
Flame Ionization Detector
Performances
Analytes and detection limit:
|
Sulphur compounds |
20 ppb |
|
Permanent gases |
10 ppm |
|
Carbon monoxide |
1 ppm |
|
Hydrocarbons |
20 ppb |
|
BTX |
20 ppb |
|
CFC |
100 ppb |
Utilities and Supplies
Gas supply (chromatography grade):
Helium 5.0 (>99.999%) - supply pressure 6 bar
Hydrogen 5.0 (>99.999%) - supply pressure 2 bar
Air 5.0 (zero grade) - supply pressure 6 bar
Instrument air or CO2 - 3-5 bar dew point -40°C
Power supply about 3KW - 230V 50Hz.
Options
Sampling and auxiliary gas control
Multiple sampling system
Automatic loading/unloading truck system under CO2 quality control
|
Front view
Dimensions
60 cm high
65 cm wide
60 cm deep.
Approximate Height
40 Kg
personal computer
excluded |  |
Automating Loading

Loading and unloading truck system under CO2 quality control
The system provides to:
• Drive plant pneumatic system to fill liquid CO2 in mobile tanks
• Analyse residuals (with CO2 analisys system)
• Analyse product filled (with CO2 analisys system)
• Print shipping documents
Truck driver follows simple instructions to connect mobile tank to plant storage.
The system atomatically transfill the product, and acquire quantities from scales and or from liquid meters.
Engineering
For CO2 gascromatographic analysis system
The system is a Shimadzu engineered GC, PC drived.
FPD and TCD detector signals are amplified with a patent Pentatechs amplifier, to improve to ppb level the sensibility of sulphur compounds and to low ppm level for permanent gases.
The system is frendly, with simple interface, alarms could be shown on the screen by colouring component box
|

| Reports can be configured with customized layouts with:
a)
Customer informations
b) Sample identification
c) Analisys results
d) Other information
|
Sampling
and auxiliary gas control
The system provides to:
- check sample and auxiliary gas pressure
- Launch alarms for pressure failures
- Switch samples to the GC system
Technical Features
Rack 19 panel
24 input
24 output
Power supply 230 V
Approx. weight 4 Kg
Communication with PC by RS232 interface
Options available for 4-20 mA signals acquisition and convertion.
Acquisition of moisture and NOx analysers signals |